Prøv at søge efter nøgleord, ordre- eller produktkode eller serienummer, f.eks. “CM442” eller “Teknisk information”
Indtast mindst 2 tegn for at starte søgningen.
Liquiline System CA80SI - silica analyzer for boiler feedwater, steam, condensate and ion exchangers

Silica analyzer
Liquiline System CA80SI

Colorimetric system for the monitoring of boiler feedwater, steam, condensate and ion exchangers

CA80SI
Please Wait (spinning wheel)
Indlæser pris
Priser er midlertidigt utilgængelige
Price on request
aktiverer pris
pris: information ønskes
from CHF ??.-
Filtrer måle- / anvendelsesopgave
Måle- / Anvendelsesopgave

Konfigurer for at downloade CAD-tegninger af dette produkt

  • Measuring principle

    Colorimetric

  • Characteristic

    Analyzer for silica in aqueous solutions

  • Measuring method

    Comply with standard colorimetric measuring principle - heteropoly blue method

  • Size

    Housing (open version):
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 in
    Housing (closed version):
    793 x 530 x 463 mm
    31.22 x 20.87 x 18.23 in
    Housing with base:
    1723 x 530 x 463 mm
    67.83 x 20.87 x 18.23 in

  • Design

    Open design / cabinet
    Plastic ASA-PC

  • Process temperature

    5 to 45 °C (41 to 113 °F)

  • Ambient temperature

    5 to 40 °C (41 to 104 °F)

  • Process pressure

    1 to 5 bar (14.5 to 72.5 psi)

  • Sample flow rate

    60 to 250 ml/min (2.03 to 8.45 fl.oz/min)

  • Consistency of the sample

    Particle free

  • Specials

    Easy upgrade to measuring station with up to four digital Memosens sensors
    Automatic calibration and cleaning
    User-configurable measuring, cleaning and calibration intervals
    optional up to 6 sample channels
    Flexible functionality and modular expandable
    Digital communication for remote access

  • Application

    Determination of silica concentration in boiler feed water
    Monitoring of water-steam-circuit and condensate
    Monitoring and optimization of reverse osmosis and desalination

  • Power supply

    100 to 120 VAC / 200 to 240 VAC ± 10%
    50 ± 1 or 60 ± 1.2 Hz

  • Output / communication

    2x 0/4 to 20 mA
    Optional: Webserver, Modbus, Ethernet/IP, Profibus DP

  • Input

    1, 2, 4 or 6 measuring channels
    Optional 1 to 4 digital inputs for sensors with Memosens protocol

  • Measurement range

    0 to 500 µg/l (ppb)
    50 to 5000 µg/l (ppb)

  • Consumables

    Reagents and standard solutions CY80SI
    Maintenance kit CAV800