Omnigrad T TST410
Modular RTD thermometer

Omnigrad T TST410
Modular RTD thermometer ©Endress+Hauser

Comprehensive and common temperature measurement technology for almost all branches of industry

The thermometer is mainly used in the chemical industry but also finds its use in other branches. Typical applications are pressure-less systems such as air ducts, flues and pipelines. For protection against high industrial demands the use of a special protective sleeve is recommended. A variety of dimensions offers flexible application possibilities.

  • Fordele

    • High degree of flexibility thanks to modular design with standard terminal heads as per DIN EN 50446 and customer-specific immersion lengths

    • High degree of insert compatibility and design as per DIN 43772

    • High accuracy and exceptional long-term stability

  • Anvendelsesområde

    Typical applications for the TST410 sensor are pressure-less systems such as air ducts, flues and pipelines.
    Special care should be given to the maximum allowed temperature surrounding the terminal head, which must be lower than 80 °C.

Funktioner og specifikationer

Tilbage

Details

  • Measuring principle

    Resistance Temperature Detector

  • Characteristic / Application

    metric style

    modular temperature assembly

    process connection as compression fitting

    without neck

    mini-head

  • Terminal head

    TA20L - aluminium - mini-head

  • Thermowell / protection tube

    without (not intended to use with thermowell)

  • Insert / probe

    mineral insulated (MI), flexible

  • Outer diameter protection tube / Insert

    3,0 mm (0,12'')

  • Max. immersion length on request

    up to 10.000,00 mm (393,70'' )

  • Material protection tube/ thermowell

    1.4401 (316)

  • Process connection

    compression fitting:

    G1/4"

  • Tip shape

    straight

  • Operating temperature range

    PT 100:

    -50 °C ...400 °C

    (-58 °F ...752 °F)

  • Max. process pressure (static)

    at 20 °C: 40 bar (580 psi)

  • Accuracy

    class A acc. to IEC 60751

    class B acc. to IEC 60751

  • Response time

    t50 = 2 s

    t90 = 4 s

  • Integration head transmitter

    no

Dokumenter / Vejledninger / Software

Tilbehør / Reservedele

Nr
Order code
Specification