Prøv at søge efter nøgleord, ordre- eller produktkode eller serienummer, f.eks. “CM442” eller “Teknisk information”
Indtast mindst 2 tegn for at starte søgningen.
Product picture OXY5500 oxygen analyzer box, right angle view

OXY5500 oxygen analyzer

Accurate and reliable O2 measurements for natural gas and gas processing applications

OXY5500
Please Wait (spinning wheel)
Indlæser pris
Priser er midlertidigt utilgængelige
Price on request
aktiverer pris
pris: information ønskes
from CHF ??.-
Filtrer måle- / anvendelsesopgave
Måle- / Anvendelsesopgave

Konfigurer for at downloade CAD-tegninger af dette produkt

  • Measuring principle

    Quenched Fluorescence

  • Product Headline

    Compact, single channel oxygen (O2) analyzer with an LCD display and built-in data logger. The sensor measures oxygen using quenched fluorescence (QF) technology, which is inserted into the process stream and is connected to the controller by an optical fiber.

  • Channels

    1

  • Analyte and Measurement ranges

    O2 (Oxygen):
    0 to 100 ppmv (minimum)
    0 to 20% volume (maximum)

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 275 kPa (20 to 40 psig)
    Sample Cell: 800 to 1200 mbara

  • Analyzer Wetted materials

    304 & 316 stainless steel
    FKM O-Rings

  • Power supply

    108 to 253 VAC, 50/60 Hz; 5.3 Watts at 120VAC; 6.6 Watts at 240VAC
    or
    9 to 30 VDC (CSA), 18 to 30 VDC (IECEx/ATEX); 4.7 Watts at 24VDC

  • Communication

    Analog Output: Two 4-20 mA outputs
    Analog Input: One 4-20 mA input (sample pressure)
    Fieldbus: RS-232C, RS-485, & Ethernet 10/100 with Modbus
    Output Relays: Qty 2, 250 mA max load (Concentration and Fault Alarms)
    USB 2.0 Works With Service Software Only
    4 GB Internal Memory with Internal Data Logging

  • Housing materials

    Electronics: 304 or 316L stainless steel

  • Hazardous area approvals

    ATEX / IECEx - Zone 2
    CSA Class I, Division 2

  • Degree of protection

    IP64, Type 4X

  • Product safety

    CE
    RCM
    FCC